Reproducible sensor sweeps and fixed-point compensation fixtures.
Note #14 · Load-Cell Telemetry
Two-Channel Differential Sampling Configuration
Sampling at 10 SPS with hardware 50/60 Hz rejection enabled yields 19.4 noise-free bits across a 2 kg full-scale aluminum beam:
sample_rate_hz = 10
pga_gain = 128
median_window = 7
ema_alpha_q15 = 4096 // 0.125 in Q1.15
Note #11 · Thermal Loop
Step-Response Summary (35 L EPP Chamber)
Measured settling time from 21.0 °C ambient to 26.5 °C target at 78% RH:
Controller Overshoot (C) 95% Settling (min) Steady RMS (C)
Bang-Bang (0.5 C) +1.35 41.0 0.44
PI + Feed-Forward +0.08 18.5 0.05
Dataset · Binary LUT v2
Reference Chamber Calibration Sweep
Raw binary snapshot containing interleaved 32-bit IEEE-754 ADC counts, die temperature readings, and polynomial compensation residuals captured over a 72-hour thermal ramp (300 KiB):
Download calibration-lut-v2.bin (300 KiB)